beautypg.com

Diagnostics and test menu, Table 3 – Hach-Lange FILTERTRAK FT 660 sc User Manual User Manual

Page 23

background image

Note: A warning can be used to trigger a relay. The sc controller can be used to
set warning levels to identify the severity of the warning. Refer to the sc controller
user manual.

Table 3 Warning list

Warning

Definition

Resolution

DARK READING

Dark reading identifies
too much light (greater
than 50 mNTU).

1. Make sure that the head

assembly is correctly
installed on the body of
the instrument.

2. From the Main menu,

select Sensor Setup,
[Select Sensor],
Calibrate, 0 Electronics.

TEMPERATURE

The internal temperature
of the head assembly is
greater than 50 °C
(122 °F).

Make sure that the operating
environment is within the
specifications. Refer to

Specifications

on page 3.

If the operating environment
is within specifications,
contact Technical Support.

5 VOLT

The monitored voltage is
outside the range 4.5–
5.5 V.

Contact Technical Support.

VOLTAGE IN

The monitored voltage is
outside the range of
9.08–14.3 V.

LASER V

The laser voltage is out
of the operating range of
4.55–5.5 V.

LASER I

The laser current is out of
operating range of 25–
75 mA.

Table 3 Warning list (continued)

Warning

Definition

Resolution

DATA LOG FULL

Only 30 minutes of
storage space is
available in the data log.

Download the log from the sc
controller to prevent loss of
data.

EVENT LOG FULL

Only 30 minutes of
storage space is
available in the event log.

OUTPUT MODE
NOT NORMAL

The instrument is being
calibrated.

Complete calibration and/or
calibration verification.

PMT VOLTAGE

The PMT voltage is out of
operating range of
11.77–15.94 V.

Contact Technical Support.

REFERENCE
VOLTAGE

The reference voltage is
out of operating range of
2.41–2.59 V.

Diagnostics and Test menu

The diagnostic and test menu shows current information about the
instrument. Refer to

Table 4

.

To access the diagnostic and test menu, select Sensor Setup, [Select
Sensor], Diag/Test.

Table 4 DIAG/TEST menu

Option

Description

SOFTWARE VERS. Shows the installed software version.

SERIAL NUMBER

Shows the serial number of the instrument.

INT TEMP

Shows the internal temperature of the instrument
electronics in °C.

DEFAULT SETUP

Sets the settings to the factory default settings. Makes the
current calibration invalid.

POWER CHECK

Shows the electrical statistics for the instrument.

English 21