Section 1 technical data – Hach-Lange RTC112 SD-Module User Manual
Page 7
7
Section 1
Technical data
These are subject to change without notice.
Embedded PC (compact industrial PC)
Processor
Pentium
®1
, MMX compatible, 500 MHz clock rate
Flash memory
2 GB compact flash card
Internal working memory
256 MB DDR-RAM (not expandable)
Interfaces
1× RJ 45 (Ethernet), 10/100 Mbit/s
Diagnostic LED
1× power, 1× LAN speed, 1× LAN activity, TC status,
1× flash access
Expansion slot
1× CompactFlash type II slot with ejector mechanism
Clock
Internal, battery-buffered clock for time and date (battery can be
replaced)
Operating system
Microsoft Windows
®2
CE or Microsoft Windows Embedded Standard
Control software
TwinCAT PLC Runtime or TwinCAT NC PTP Runtime
System bus
16 bit ISA (PC/104 standard)
Power supply
Via system bus (through power supply module CX1100-0002)
Max. power loss
6 W (including the system interfaces CX1010-N0xx)
Analog inputs
0/4 to 20 mA for input of the feed flow rate and the polymer flow rate
Number of inputs
One-channel: 2 (KL3011)
Two-channel: 4 (KL3011)
Internal resistance
80 ohm + diode voltage 0.7 V
Signal current
0/4 to 20 mA
Common mode voltage (U
CM
)
35 V max.
Measurement error (for entire measurement
range)
< ± 0.3 % (from end value of measurement range)
Voltage surge resistance
35 V DC
Electrical isolation
500 V
eff
(K-bus/signal voltage)
Analog outputs
Output of the polymer dosing, output of the feed flow rate
Number of outputs
One-channel: 2 (KL4012)
Two-channel: 4 (KL4012)
Supply voltage
24 V DC via the power contacts
(Alternatively, 15 V DC with bus termination KL9515)
Signal current
0/4 to 20 mA
Working resistance
<500 ohm
Measurement error
± 0.5 LSB linearity error
± 0.5 LSB offset error
± 0.1 % (relative to the measuring range end value)
Resolution
12 bit
Conversion time
Approximately 1.5 ms
Electrical isolation
500 V
eff
(K-bus/field voltage)