Section 8 troubleshooting – Hach-Lange Piston pipette User Manual

Page 27

background image

25

Section 8

Troubleshooting

The following table is intended to help you to identify and
eliminate errors:

Table 2 Troubleshooting

Error

Possible cause

Response

Leakage

Tip not attached
properly

Fix tip firmly in place

Foreign bodies
between cone and
tip

Clean the cone and
attach a new tip

Foreign bodies
between O-ring and
piston

Clean and grease
the O-rings and the
piston in the cylinder.
Use the supplied
grease

Insufficient grease
on the O-rings and
the piston

Grease adequately

Defective O-rings

Replace O-rings

Inaccurate
pipetting

Incorrect operation

Use pipette as
described in the
manual

Tip not firmly
attached

Attach tip firmly

Calibration has
changed, e.g. due to
faulty operation

Recalibrate