Section 6 calibration, 1 checking the calibration – Hach-Lange Piston pipette User Manual
Page 24

22
Section 6
Calibration
The pipette is calibrated in the factory with distilled water at
22 °C/72 °F. No recalibration is needed for routine use. The
pipette can, however, be recalibrated for other temperatures
and solutions of different viscosities. An appropriately
sensitive analytical balance, a small beaker and distilled water
are needed to carry out the calibration.
Important Note: If either the handle or the piston is replaced,
the pipette must be recalibrated. If the O-rings or the tip cone
are replaced, the calibration should be checked.
6.1 Checking the calibration
1. Set the pipette volume as follows in
Table 1 Checking the
calibration
:
Table 1 Checking the calibration
Pipette
(µL)
Volume setting
permissible
weight range
(mg)
10
10
9.8–10.2
20 20
19.8–20.2
50
50
49.5–50.5
100 100
99.2–100.8
200
200
198.7–201.3
500 500
497.0–503.0
1000
1000
994.0–1006.0
2000 2000
1990.0–2010.0
200–1000
300
298–302
1000–5000 2000 1990–2010
AV_pipette.book Seite 22 Donnerstag, 8. Dezember 2005 3:56 15
- AMTAX sc (118 pages)
- PHOSPHAX sc (106 pages)
- CL17 Instruction sheet (8 pages)
- CL17 USER MANUAL (60 pages)
- CLF10 sc (382 pages)
- 9184 sc (50 pages)
- POLYMETRON 9611 sc Installation (398 pages)
- POLYMETRON 9610 sc Maintenance and Troubleshooting (286 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SEQUENCER LINE INSTALLATION (2 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SS SAMPLE CONDITIONING (4 pages)
- HACH 5500 sc Installation (408 pages)
- HACH 5500 sc Maintenance and Troubleshooting (286 pages)
- HACH 5500 sc PO43-HR Operations (390 pages)
- HACH 5500 sc PO43-LR Operations (392 pages)
- HACH 5500 sc SiO2 Installation (418 pages)
- HACH 5500 sc SiO2 Operations (394 pages)
- POLYMETRON 9610 sc SiO2 Installation (410 pages)
- POLYMETRON 9240 Basic User Manual (162 pages)
- POLYMETRON 9240 Operator Manual (144 pages)
- PHOSPHAX sigma Basic User Manual (305 pages)
- AMTAX inter2 Basic User Manual (377 pages)
- GANIMEDE P (60 pages)
- BODTrak II Basic User Manual (172 pages)
- BODTRACK II User Manual (28 pages)
- BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) (46 pages)
- SPECTRO COLOR d-8 (29 pages)
- REFO 60_REFO 60D (36 pages)
- REFO 3_REFO 3D (19 pages)
- LICO 620 Basic User Manual (329 pages)
- LICO 620 User Manual (114 pages)
- SD900 (2 pages)
- LZX971 (8 pages)
- SC Sensor Plug and SC Sensor Coupling (8 pages)
- HQ Series Portable Meters (4 pages)
- LDO Sensor Replacement Kit for use with Hach HQ10 and HQ20 (43 pages)
- LDO Sensor Replacement Kit (8 pages)
- SIP 10 (273 pages)
- SC 100 (58 pages)
- SI792 E_T Quick start guide (2 pages)
- SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual (116 pages)
- SI792 D Quick start guide (2 pages)
- SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual (98 pages)
- SI794 C and SI794 D Short instructions (2 pages)
- SI794 D tr User Manual (108 pages)
- SI792 C Quick start guide (2 pages)