3 [+devicedata] menu, Operation – Hach-Lange FILTRAX User Manual User Manual
Page 38

36
Operation
5.3 [+DEVICEDATA] menu
Instrument parameters for adaptation to the related conditions.
Note: While the instrument is in this menu, sample pumping continues. If a key is not
pressed for longer than ten minutes, the instrument automatically returns to the display of
the data on operation.
*If the power supply is interrupted (power failure) both relay contacts are closed.
MENU
Command
Setting
Description
+DEVICEDATA
CONTRAST
-90 to +90
Reading angle for the display
LANGUAGE
D, GB, NL, F, I, E, PL, S
HEATER
ON, OFF, TIMER (time control)
Default setting: TIMER
Operation of the tube heaters
HEAT.START (only for [HEATER]:TIMER)
JAN to DEC,
Default setting: OCT
Activation of the heater at the start of the month selected
HEAT.STOP (only for [HEATER]:TIMER)
JAN to DEC,
Default setting: MRZ (March)
Deactivation of the heater at the end of the month selected
WARNING<
200...800 ml/h in 10 ml steps
Warning relay contact* set when the sample volume drops below
the amount set
FAULT<
100...600 ml/h in 10 ml steps
Fault relay contact* set when the sample volume drops below the
amount set
DATE
Setting for the date
TIME
Setting for the time
PASSWORD
4-digits with numbers 1-4
Password protection on activation
SW-VERSION
Only for information
Version of the instrument software
SW-CO.PROC.
Only for information
Version of the coprocessor software