Section 5 maintenance, 1 maintenance schedule, 2 clean the sensor – Hach-Lange NH4D sc Ammonium Sensor User Manual
Page 29: 3 replace the sensor cartridge

29
Section 5
Maintenance
Important Note: Only qualified personnel should conduct the tasks 
described in this section of the manual.
5.2 Clean the sensor
Important Note: 
Do not touch the membranes with fingers.
Do not clean the sensor cartridge with sharp objects that may 
cause scratches and do not use any chemical cleaning agent.
1.
Clean the sensor cartridge with the soft brush supplied.
2.
Clean the sensor body (not the cartridge) with a sponge or 
brush.
3.
Rinse the sensor with clean, lukewarm tap water.
Note: Do not use destilled water.
5.3 Replace the sensor cartridge
The sensor cartridge is replaced as described below and in
Figure 12 on page 30
.
1.
Clean the sensor and thoroughly dry the sensor cartridge and 
the sensor adapter.
2.
Unscrew the locking ring and remove.
Important Note: The sensor cartridge must direct downwards so 
that no water can run into the sensor adapter. The contacts 
between the sensor and the sensor cartridge must remain dry.
3.
Pull the sensor cartridge out of the sensor adapter and dispose 
of the old sensor cartridge and opaque gasket per applicable 
regulations.
4.
Ensure that the new opaque gasket is installed during every 
sensor cartridge replacement.
5.1 Maintenance schedule
Maintenance task
30 days
1
12 months
Clean sensor
2
x
Replace the sensor cartridge
3,
4
x
Check sensor for damage
x
Check measured value by lab reference analysis and correct values by matrix 
correction if required
3
x
1
Recommendation: weekly during the first month in operation
2
The frequency of cleaning depends on the application. In some applications more or less frequent cleaning may be necessary.
3
In typical operating conditions, a different interval may be required depending on the specific application and local conditions.
4
Sensor cartridges are wear parts and are not covered by the instrument warranty.
Important Note: Do not check the sensor with usual NH
4
–N standard solutions, because the ion strength of
normal standard solutions are not strong enough. Use therefore calibration solutions with artificial ion matrix 
generate to have plausible measurements (e.g. LCW892).
