4.6.2 Matrix correction overview
With the NH4D sc sensor, there are various options (
Table 1
) for
correcting the measured value with laboratory reference values.
For the laboratory reference value, a sample is taken for laboratory 
analysis and the measurement value at that time is stored.
Depending on the parameter set, the laboratory reference value of 
the water sample is stored as ammonium nitrogen (NH
4
–N) or as
ammonium (NH
4
). This laboratory reference value replaces the
prior value measured by the sensor.
Table 1 Correction options for the NH4D sc sensor
Correction Option
Application
MATX1
The most commonly used Matrix correction. It performs a 1 point matrix correction (offset 
adjustment) for ammonium (
).
MATX1 K+
With the MATX1 K+ the ammonium value will be corrected as with the MATX1 and in addition the 
potassium value can be corrected. This is necessary if a higher precision of the ammonium value is 
required as the potassium value will interfere the ammonium measurement (
MATX2
Where there is a dynamic process with a large ammonium fluctuation (at least 1/2 decade between 
the lowest and highest concentration)
1
, it is recommende to perform the MATX2 (
MATX2 K+
Where there is a dynamic process with a large ammonium fluctuation (at least 1/2 decade between 
the lowest and highest concentration)
1
and potassium value is to be corrected, it is recommended to
perform the MATX2 K+ (
).
VALUE CORR
The ammonium and potassium displayed value and the ammonium laboratory value can be entered 
for two points. This is a different way to perform a MATX2 Correction. Here, for two points, the 
displayed ammonium and potassium values and the ammonium laboratory value are entered via an 
input screen.
PREVIOUS CAL
Activation of one of the last four matrix and value corrections performed.
FACTORY CAL
If the current sensor code is no longer available, average data for the sensor can be activated using 
FACTORY CAL.
1
Example of a half decade: The ammonium nitrogen concentration moves between 1 and 5 mg NH
4
–N or
between 5 and 25 mg/l NH
4
–N.