Section 1 specifications – Hach-Lange UVAS sc User Manual

Page 5

background image

5

Section 1

Specifications

Specifications are subject to change without notice.

Table 1 UVAS plus sc tank sensors

Table 2 UVAS plus sc bypass sensors

UVAS plus sc

Measuring technique

UV absorption measurement (2-beam technique), reagent-free

Measuring method

SAC 254 in accordance with DIN 38404 C3

Measuring path

1, 2, 5 and 50 mm

Measuring range

0.01-60 m

–1

(50 mm)

0.1-600 m

–1

(5 mm)

0-1500 m

–1

(2 mm)

2-3000 m

–1

(1 mm)

Can be calibrated to the total parameter COD depending on the application

Compensation

550 nm

Measuring interval (> min)

>1 min

Cable length

10 m (33 ft) standard

Optional extension cables available in 5, 10, 15, 20, 30, 50 m

Total maximum length: 60 m (196 ft)

Control function

PID, time control, 2-point controller (with sc 100)

Probe pressure limit

max. 0.5 bar

Ambient temperature

+2 °C to +40 °C

Dimensions D × L

Approx. 70 mm x 333 mm

Mass

approx. 3.6 kg

Inspection interval

6 months

User maintenance

1 h / month, typical

UVAS plus sc (in the bypass)

Measuring path

2, 5 and 50 mm

Measuring range
With NO

3

-N standard solutions

0.01–60 m

-1

(50 mm)

0.1–600 m

-1

(5 mm)

0–1500 m

-1

(2 mm)

Can be calibrated to the total parameter COD depending on the application

Cable length

10 m (33 ft) standard

Optional extension cables available in 5, 10, 15, 20, 30, 50 m

Total maximum length: 60 m (196 ft)

Sample flow rate

At least 0.5 L/h sample

Pressure limit

max. 0.5 bar

Sample connection

Hose ID 4 mm / OD 6 mm

Sample temperature

+2 °C to +40 °C

Dimensions

see

Figure 4, page 14