Section 6

Troubleshooting

6.1 Error messages

Possible sensor error messages are displayed by the sc controller.

6.2 Warnings

Possible sensor warning messages are displayed by the sc controller.

Table 4 Error messages

Error displayed

Rectification

NONE

MOIST

Check MOIST value on the SENSOR SETUP menu,
TEST/MAINT, MAINT.PROC., SIGNALS.
Remove sensor from the tank and call service

R < M

Call service

DEXT < 0.0

Check calibration, call service

W.POS. UNKNOWN

Check measuring path, call service

W. BLOCKED

Check measuring path, call service

FLASH FAILURE

Call service

R TOO HIGH

Call service

Table 5 Warnings

Warning displayed

Cause

Rectification

NONE

Correct measuring operation

EM TOO HIGH

Turbidity, organic content or nitrate concentration too
high, measuring range exceeded as a result

Check measurement in the laboratory

CONC. TOO HIGH

Nitrate concentration too high, as a result measuring
range exceeded

Check measurement in the laboratory

CHECK KALIBR.

Test interval elapsed

Check calibration

REPLACE PROFILE

Counter elapsed

Change wiper profile

SERVICE REQUIRED

Counter elapsed

Call service

REPLACE SEALS

Counter elapsed

Call service

SHAFT SEALS REPL.

Counter elapsed

Call service