Section 6 troubleshooting, 1 error messages, 2 warnings – Hach-Lange UVAS sc User Manual
Page 31: 1 error messages 6.2 warnings
31
Section 6
Troubleshooting
6.1 Error messages
Possible sensor error messages are displayed by the sc controller.
6.2 Warnings
Possible sensor warning messages are displayed by the sc controller.
Table 4 Error messages
Error displayed
Rectification
NONE
MOIST
Check MOIST value on the SENSOR SETUP menu,
TEST/MAINT, MAINT.PROC., SIGNALS.
Remove sensor from the tank and call service
R < M
Call service
DEXT < 0.0
Check calibration, call service
W.POS. UNKNOWN
Check measuring path, call service
W. BLOCKED
Check measuring path, call service
FLASH FAILURE
Call service
R TOO HIGH
Call service
Table 5 Warnings
Warning displayed
Cause
Rectification
NONE
Correct measuring operation
EM TOO HIGH
Turbidity, organic content or nitrate concentration too
high, measuring range exceeded as a result
Check measurement in the laboratory
CONC. TOO HIGH
Nitrate concentration too high, as a result measuring
range exceeded
Check measurement in the laboratory
CHECK KALIBR.
Test interval elapsed
Check calibration
REPLACE PROFILE
Counter elapsed
Change wiper profile
SERVICE REQUIRED
Counter elapsed
Call service
REPLACE SEALS
Counter elapsed
Call service
SHAFT SEALS REPL.
Counter elapsed
Call service