Two- and three-point calibration, Operation – Hach-Lange SENSION+ EC5 User Manual
Page 16

12
6. Operation
6.3.3. Two- and three-point calibration
Follow the diagram below.
The equipment will repeat the process described for the first stan-
dard solution with the second and third.
Dip the sensor into the tube several times to homogenize the liq-
uid. Do not hold it by the tube to avoid heating of the standard
solution.
Note: if the measuring instrument is not used for 5 minutes, it shuts
off automatically.
If errors occur during calibration, an error message will appear
on screen (see pg. 19).
When you switch standards, rinse the cell with deionized water.
μS/cm
μS/cm
ºC
T
t
μS/cm
EC
ºC
T
t
μS/cm
EC
μS/cm
ºC
T
t
μS/cm
EC
ºC
T
t
μS/cm
EC
@25ºC
μS/cm
ºC
EC
T
t
50 60
0.2... 200.000 μS/cm -30... 80
ºC
50 60
0.2... 200.000 μS/cm -30... 80
ºC
50 60
0.2... 200.000 μS/cm -30... 80
ºC
50 60
0 60
6
0
2
200
000
S/c
0.2... 200.000 μS/c
m3
0
8
m -30... 8
0
0
ºC
ºC
0
0
0
50 60
0 60
6
0
2
200
000
S/
0.2... 200.000 μS/c
30
8
m -30... 8
0
0
ºC
ºC
50 60
50 60
0.2... 200.000 μS/cm -30... 80
ºC
50 60
50 60
0 60
6
0
0
0
50 60
0 60
6
Screw the
sensor
to the 1st tube
Screw the sen-
sor to the 2nd
tube
To exit calibration
Two-point calibration
To exit calibration.
One-point calibration
sensION+ EC5 HACH_LANGE 0710.indd ANG:12
sensION+ EC5 HACH_LANGE 0710.indd ANG:12
6/10/10 12:03:21
6/10/10 12:03:21