Service timer, Language selection, Clock – Hach-Lange ORBISPHERE KM1100 Basic User Manual User Manual

Page 22: Other menus, Maintenance, Instrument maintenance, Sensor maintenance, Equipment required

background image

• The display shows the current instrument date and time, next calibration due date and time, and

the remaining days.

The next calibration date is updated when the sensor is calibrated.

Service timer

The sensor attached to your instrument will require periodic servicing and maintenance. The
instrument can automatically remind the user when the next sensor service is due.

• To enable the timer, select Enable and enter a delay in days.
• The display shows the current instrument date and time, the next sensor service due date and

time, and the remaining days.

After a service press the Service done button to update the next service date.

Language selection

Select the language from a list and restart the instrument to apply the change.

Clock

Update the time and date information.

Other menus

For information on setting up relays and analog outputs refer to the full user manual (Inputs/Outputs
menu).
For information on setting up RS485, PROFIBUS-DP, USB, HTTP/TCP-IP and attached PRINTER
links refer to the full user manual (Communications menu).
For information on setting up products and global configurations refer to the full user manual
(Products and Global Configuration menus).

Maintenance

Instrument maintenance

C A U T I O N

Personal Injury Hazard. Any instrument maintenance should be carried out by a qualified Hach Lange Service
Technician. Please contact your local representative should you feel any maintenance or instrument adjustments
are required.

Sensor maintenance

The sensor spot needs to be replaced about once a year. The procedure is very simple and takes no
more than a couple of minutes.

Equipment required

Note: Check the Ksv factor of the new spot on the maintenance kit box. If this is different to the Ksv value displayed
in the factory parameter (see

Configure manual calibration

on page 20) update the factory parameter to the new

value.

1. A replacement sensor spot
2. The maintenance tool delivered with the sensor
3. O-ring supplied with the sensor spot

22 English