Hach-Lange SI794 D User Manual User Manual
Page 93
Cleaning function *
Relay contact, floating,
for controlling a simple rinsing system
Contact ratings
AC < 250 V / < 3 A / < 750 VA
DC < 30 V / < 3 A / < 90 W
Contact response *
N/C or N/O
Interval *
000.0 ... 999.9 h
(000.0 h = cleaning function switched off)
Rinse duration *
0000 ... 1999 s
Display
LC display, 7-segment with icons
Main display
Character height 17 mm, unit symbols 10 mm
Secondary display
Character height 10 mm, unit symbols 7 mm
Sensoface
3 status indicators (friendly, neutral, sad face)
Mode indication
5 mode indicators
"meas", "cal", "alarm", "cleaning", "config"
18 further icons for configuration and messages
Alarm indication
Red alarm LED in case of alarm or HOLD (user defined)
Keypad
5 keys
Service functions
Current source
Current specifiable for output 1 and 2 (00.00 to 22.00 mA)
Manual controller
Controller output entered directly (start of control process)
Device self-test
Automatic memory test (RAM, FLASH, EEPROM)
Display test
Display of all segments
Last Error
Display of last error occurred
Sensor monitor
Display of direct, uncorrected sensor signal
Relay test
Manual control of the four switching contacts
Parameter sets
Two selectable parameter sets for different applications
Data retention
Parameters and calibration data > 10 years (EEPROM)
93