Specifications – Hach-Lange SI794 D User Manual User Manual
Page 92
Alarm contact
Relay contact, floating
Contact rating
AC < 250 V / < 3 A / < 750 VA
DC < 30 V / < 3 A / < 90 W
Contact response
N/C (fail-safe type)
Response delay *
0000 ... 0600 sec
Limit values
Output via relay contacts R1, R2
Contacts R1, R2, floating but inter-connected
Contact rating
AC < 250 V / < 3 A / < 750 VA
DC < 30 V / < 3 A / < 90 W
Contact response *
N/C or N/O
Response delay *
0000 ... 9999 s
Setpoints *
Within selected range
Hysteresis *
000.0 ... 050.0 % / 00.00 ... 05.00 mg/L (ppm)
PID process controller Output via relay contacts R1, R2
(Relay R1: below setpoint,
Relay R2: above setpoint)
Setpoint specification * 0 ... 500 % / 0.00 ... 50.00 mg/L (ppm)
Neutral zone *
000.0 ... 050.0 % / 00.00 ... 05.00 mg/L (ppm)
P-action component * Controller gain Kp:
0010 ... 9999 %
I-action component * Reset time Tr:
0000 ... 9999 s
(0000 s = no integral action)
D-action component * Rate time Td:
0000 ... 9999 s
(0000 s = no derivative action)
Controller type *
Pulse length controller or pulse frequency controller
Pulse period *
0001 ... 0600 s, min. ON time 0.5 s
(pulse length controller)
Max. pulse frequency * 0001 ... 0180 min
-1
(pulse frequency controller)
Specifications
92