Default settings of parameter sets – Hach-Lange SI794 D User Manual User Manual
Page 61
![background image](/manuals/332847/61/background.png)
Two complete parameter sets are stored in the EEPROM.
As delivered, the two sets are identical but can be edited.
Note:
Fill in your configuration data on the following pages.
Code. Parameter
Default setting
o1. Sensor type
A
o1. %, mg/L, ppm
%
o1. 0/4-20 mA
4-20 mA
o1. Current start
0000 %
o1. Current end
0500 %
o1. Filter time
0 sec
o1. 22mA signal
OFF
o1. HOLD response
Last
o1. Fix current
021.0 mA
o2. Unit °C / °F
°C
o2. Temp probe
22 NTC
o2. 0/4 ...20mA
4-20 mA
o2. Current start
000.0 °C
o2. Current end
100.0 °C
o2. Filter time
0 sec
o2. 22mA signal
OFF
o2. HOLD response
Last
o2. Fix current
021.0 mA
Co. Polariz. voltage
675 mV
Co. Pressure unit
bar
Co. Pressure
1.013 bar
Co. Salinity
00.00 mg/L
CA. Cal mode
Sat
CA. Cal interval
0000 h
AL. Sensocheck
OFF
AL. Alarm delay
0010 sec
AL. LED Hold
OFF
Code. Parameter
Default setting
rL. Relay function
Limit
L1. Contact function
Lo
L1. Contact response
N/C
L1. Setpoint
0000 %
L1. Hysteresis
0001 %
L1. Delay
0010 sec
L2. Contact function
Hi
L2. Contact response
N/C
L2. Setpoint
0500 %
L2. Hysteresis
0001 %
L2. Delay
0010 sec
Ct. Setpoint
0100 %
Ct. Neutral zone
0010 %
Ct. P action
0100 %
Ct. I action
0000 sec
Ct. D action
0000 sec
Ct. PLC/PFC controller
PLC
Ct. Pulse length
0010 sec
Ct. Pulse frequency
0060 /min
Ct. Hold response
Last
Cn. Rinse/Par.set 2
rinse
Cn. Rinsing interval
000.0 h
Cn. Rinse time
0060 sec
Cn. Contact type
N/C
Default settings of parameter sets