On piping, In a bypass, Figure 2 – Hach-Lange POLYMETRON 9500 Contacting Conductivity Sensors User Manual User Manual
Page 6: Figure 3

• B: Good - satisfactory immersion of the electrode surfaces and good sample flow.
• C: Poor - incomplete immersion of the electrodes or incorrect sample flow.
On piping
Immerse the internal electrode completely in the process sample.
Note: In the following illustrations, the arrows indicate the sample flow direction.
Figure 2 Model 8315 sensor mounting example
Figure 3 Model 8394 sensor mounting example
The 8394 sensor installs perfectly in a Tri-Clover
®
Tri-Clamp
™
Tee starting from 1.5 inch diameter
(A), and also at a 90° angle starting from 2 inches (B). All Tri-Clamp
™
Tees are in compliance with
3A standards for CIP (cleaning In place).
In a bypass
Polymetron flow chambers are designed not to retain air bubbles. To facilitate the extraction of
bubbles, use a minimum flow rate of 20 L/h (ideally 60 L/h).
Note: The progressive accumulation of bubbles on the surface of the sensor reduces the active surface, increases
the cell constant, and leads to an abnormally low conductivity measurement.
6 English
- AMTAX sc (118 pages)
 - PHOSPHAX sc (106 pages)
 - CL17 Instruction sheet (8 pages)
 - CL17 USER MANUAL (60 pages)
 - CLF10 sc (382 pages)
 - 9184 sc (50 pages)
 - POLYMETRON 9611 sc Installation (398 pages)
 - POLYMETRON 9610 sc Maintenance and Troubleshooting (286 pages)
 - POLYMETRON 9611 sc USER INSTRUCTIONS: SEQUENCER LINE INSTALLATION (2 pages)
 - POLYMETRON 9611 sc USER INSTRUCTIONS: SS SAMPLE CONDITIONING (4 pages)
 - HACH 5500 sc Installation (408 pages)
 - HACH 5500 sc Maintenance and Troubleshooting (286 pages)
 - HACH 5500 sc PO43-HR Operations (390 pages)
 - HACH 5500 sc PO43-LR Operations (392 pages)
 - HACH 5500 sc SiO2 Installation (418 pages)
 - HACH 5500 sc SiO2 Operations (394 pages)
 - POLYMETRON 9610 sc SiO2 Installation (410 pages)
 - POLYMETRON 9240 Basic User Manual (162 pages)
 - POLYMETRON 9240 Operator Manual (144 pages)
 - PHOSPHAX sigma Basic User Manual (305 pages)
 - AMTAX inter2 Basic User Manual (377 pages)
 - GANIMEDE P (60 pages)
 - BODTrak II Basic User Manual (172 pages)
 - BODTRACK II User Manual (28 pages)
 - BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) (46 pages)
 - SPECTRO COLOR d-8 (29 pages)
 - REFO 60_REFO 60D (36 pages)
 - REFO 3_REFO 3D (19 pages)
 - LICO 620 Basic User Manual (329 pages)
 - LICO 620 User Manual (114 pages)
 - SD900 (2 pages)
 - LZX971 (8 pages)
 - SC Sensor Plug and SC Sensor Coupling (8 pages)
 - HQ Series Portable Meters (4 pages)
 - LDO Sensor Replacement Kit for use with Hach HQ10 and HQ20 (43 pages)
 - LDO Sensor Replacement Kit (8 pages)
 - SIP 10 (273 pages)
 - SC 100 (58 pages)
 - SI792 E_T Quick start guide (2 pages)
 - SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual (116 pages)
 - SI792 D Quick start guide (2 pages)
 - SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual (98 pages)
 - SI794 C and SI794 D Short instructions (2 pages)
 - SI794 D tr User Manual (108 pages)
 - SI792 C Quick start guide (2 pages)
 