Specifications – Hach-Lange SI794 C User Manual User Manual
Page 102

Display
LC display, 7-segment with icons
Main display
Character height 17 mm, unit symbols 10 mm
Secondary display
Character height 10 mm, unit symbols 7 mm
Sensoface
3 status indicators (friendly, neutral, sad face)
Mode indication
5 mode indicators "meas", "cal", "alarm", "cleaning", "config"
18 further icons for configuration and messages
Alarm indication
Red LED in case of alarm or HOLD, user defined
Keypad
5 keys: [cal] [conf] [
4
] [
5
] [enter]
Service functions
Current source
Current specifiable for output 1 and 2 (00.00 to 22.00mA)
Manual controller
Controller output entered directly (start of control process)
Device self-test
Automatic memory test (RAM, FLASH, EEPROM)
Display test
Display of all segments
Last Error
Display of last error occurred
Sensor monitor
Display of direct sensor signal (resistance/temperature)
Relay test
Manual control of the four switching contacts
Parameter sets
*
Two selectable parameter sets for different process phases
Switchover via CONTROL input or manually
Signaling via relay contact PSet2
Data retention
Parameters and calibration data > 10 years (EEPROM)
EMC
EN 61326
Emitted interference
Class B (residential area)
Class A for mains > 60 V DC
Immunity to interference Industry
102
Specifications