Specifications – Hach-Lange SI794 D tr User Manual User Manual
Page 90
O
2
input
Type A sensor:
Hamilton Oxyferm, Oxysens
Type B sensor:
Hamilton Oxygold
Measuring current
0 ... 600 nA
Resolution
10 pA
Saturation (-10 ... 80 °C)
0... 120 %
Measurement error
1,2,3
0.5 % meas.val. + 0.1 %
Concentration (-10 ... 80 °C) 0000 ... 9999 µg/L
0000 ... 9999 ppb
0.000 ... 9.999 mg/L
0.000 ... 9.999 ppm
0.00 ... 50.00 ppm
Measurement error
1,2,3
0.5 % meas.val. + 0.005 mg/L or 0.005 ppm
Permissible guard current
≤ 20 µA
Polarization voltage *
0 ... 1000 mV, resolution 3 mV
Process pressure *
0 ... 9.999 bars
( ... 999.9 kPa / ... 145.0 PSI)
Salinity correction *
00.00 ... 45.00 g/kg
Sensor standardization
Operating modes *
O
2
saturation (automatic)
O
2
concentration (automatic)
Product calibration
Zero calibration
Calibration range
Zero point
± 2 nA
Sensor Type A
Slope
25 ... 130 nA
(at 25 °C, 1013 mbars)
Calibration range
Zero point
± 2 nA
Sensor Type B
Slope
200 ... 550 nA
(at 25 °C, 1013 mbars)
Calibration timer *
0000 ... 9999 h
Pressure correction *
0.000 ... 9.999 bars / 999.9 kPa / 145.0 PSI
Sensocheck
Monitoring for short circuits / open circuits
(can be disabled)
Sensoface
Provides information on the sensor condition
evaluation of zero point/slope, response time,
calibration interval, Sensocheck
Specifications