Specifications – Hach-Lange SI794 D tr User Manual User Manual

Page 90

background image

O

2

input

Type A sensor:

Hamilton Oxyferm, Oxysens

Type B sensor:

Hamilton Oxygold

Measuring current

0 ... 600 nA

Resolution

10 pA

Saturation (-10 ... 80 °C)

0... 120 %

Measurement error

1,2,3

0.5 % meas.val. + 0.1 %

Concentration (-10 ... 80 °C) 0000 ... 9999 µg/L

0000 ... 9999 ppb

0.000 ... 9.999 mg/L

0.000 ... 9.999 ppm

0.00 ... 50.00 ppm

Measurement error

1,2,3

0.5 % meas.val. + 0.005 mg/L or 0.005 ppm

Permissible guard current

≤ 20 µA

Polarization voltage *

0 ... 1000 mV, resolution 3 mV

Process pressure *

0 ... 9.999 bars

( ... 999.9 kPa / ... 145.0 PSI)

Salinity correction *

00.00 ... 45.00 g/kg

Sensor standardization

Operating modes *

O

2

saturation (automatic)

O

2

concentration (automatic)

Product calibration

Zero calibration

Calibration range

Zero point

± 2 nA

Sensor Type A

Slope

25 ... 130 nA

(at 25 °C, 1013 mbars)

Calibration range

Zero point

± 2 nA

Sensor Type B

Slope

200 ... 550 nA

(at 25 °C, 1013 mbars)

Calibration timer *

0000 ... 9999 h

Pressure correction *

0.000 ... 9.999 bars / 999.9 kPa / 145.0 PSI

Sensocheck

Monitoring for short circuits / open circuits

(can be disabled)

Sensoface

Provides information on the sensor condition

evaluation of zero point/slope, response time,

calibration interval, Sensocheck

Specifications

90