Alarm contact

Cleaning contact

LED

Timeout

Operating

status

Measuring

Operating states

Rel.1/2 Controller

Rel.1/2 Limit value

Out 2

Out 1

Cal Info

(cal) 0000

20

s

Error Info

(conf) 0000

20

s

Calibration

(cal) 1100

Temp adjustment

(cal) 1015

Calibration error messages

Slope out of range

Wrong calibration values specified (relative

humidity, pressure, saturation, concentration)

Icon blinks:

Problem

Possible causes

Calibration aborted after 12 minutes

Sensor defective or dirty

• No electrolyte in the sensor

• Sensor cable insufficiently shielded or defective

• Strong electric fields influence the measurement

• Temperature fluctuation of calibration solution

In addition

“CAL Err”

flashes.