The arrows indicate

which relay (valve)

is active:

Controller

characteristic

+100 %

-100 %

Relay 2 active

(Meas. value >

setpoint)

Relay 1 active

(Meas. value <

setpoint)

Setpoint

Controller

output

Momentary controller output (adjusted value has

not been saved yet)

Controller test (manual specification of controller

output)

• Press conf, enter code: 5559

After function activation “Ctrl” is displayed for

approx. 3 sec. With the controller turned off, “OFF” is

displayed in addition, then return to measuring

mode.

The function is used to start up control loops or

check the actuators.

For bumpless changeover to automatic operation

(exiting this function), configure an I-action

component (reset time).

Specify value:

Select usingkey, edit number usingkey.

Press enter to proceed.

The transmitter is in Hold mode.

Press enter to return to measurement (Hold remains

active for another 20 sec).

Controller output -100 to 0 %: Relay 2 active

Controller output 0 to +100 %: Relay 1 active

Display

Action / Remarks