beautypg.com

2 calibration in water, 3 gas calibration (volume concentration), Calibration – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual

Page 76

background image

Calibration

74

8.1.2.2 Calibration in water
The water must be in equilibrium with air (percent saturation for water is
100%). The exchange between air and water is very slow. To speed up
the calibration, make sure that there is a steady flow during calibration.
1. Press

CAL

, enter passcode: 1100,

ENTER

. The display will briefly

show CAL and then the entered (or default) humidity value.

2. Remove the sensor from the process and place in the calibration

water.

3. Enter the actual relative humidity (for water always 100% relH) and

press

ENTER

. The display will show the entered (or default)

barometric pressure.

4. Enter the barometric pressure and press

ENTER

. A drift check will

begin. The drift check may take some time. When complete, the
display will show the concentration in the units specified in the
configuration menu.
Note: To exit the drift check after > 10 seconds, press

CAL

(accuracy will be reduced).

5. Accept the calculated value for the concentration or adjust the value

as necessary and then press

ENTER

. The display will show the new

slope and zero (relative to 25 °C and 1.013 mbar).

6. Press

ENTER

to end the calibration. Return the sensor to the

process. The measured value will be shown in the display
alternately with Hold.

7. Press

ENTER

to return to the measure mode.

8.1.3

Gas calibration (volume concentration)

Make sure that GAS is selected in the configuration menu (

Table 9 on

page 36

).

1. Press

CAL

, enter passcode: 1100,

ENTER

. The display will briefly

show CAL and then the entered (or default) humidity value.