beautypg.com

Operation—foundation fieldbus, 2 configuration menu (continued) – Hach-Lange SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual User Manual

Page 47

background image

Operation—Foundation Fieldbus

45

Co.PrES

0.000–9.999 BAR (default: 1.013 BAR)
999.9 kPa / 145.0 PSI (not in sensor)

Select salinity correction
Co.SAL

0.00–45.00 mg/L (default: 0.00 mg/L)

Select calibration mode

Ca.MOD

SAt (default)
Conc

Select calibration interval
CA.tiME

0000–9999 h (default 0000 h)

Select Sensocheck

AL.SnSO

CHECK ON

Continuous Sensocheck evaluation of
sensor function

CHECK OFF (default)

LED in Hold mode

AL.LED

HOLD ON

LED blinks during hold

HOLD OFF (default)

LED off during hold

Select Fieldbus address (optional)

1

FF.ADR

0017–0031 BUS (default: 0026 BUS)

1

Use only when there is no bus connection. The transmitter will restart and set all

parameters to default values. Individual settings must be entered once more.

6.1.2

Configuration menu (continued)