Operation—si792(x) D (4–20 mA/HART)

41

5.2.4 Calibration mode configuration

An overview of the calibration setup menu is shown in

Table 12

. Refer

to

Appendix A on page 95

for buffer tables.

5.2.5 Alarm settings configuration

An overview of the alarm setup menu is shown in

Table 13

.

Table 12 Calibration mode setup menu—CAL

MNU

Select calibration units

CA. MOD

SAt (default)
Conc

Enter calibration timer interval
CA.tiME

Enter the time interval for calibration (0 to 9999 hours; default: 0 h)

Table 13 Output setup menu for alarm settings—ALrt

MNU

Select Sensocheck on or off

AL.SnSO

CHECK ON

Continuous evaluation of sensor function

CHECK OFF (default)

Enter alarm delay
AL.dLY

0010 sec (default)

Range: 0000–0600 sec

LED in Hold mode

AL.LED

HOLD ON

LED blinks during hold

HOLD OFF

LED off during hold