beautypg.com

Prognosys indicator bars, Figure 3, For the measurement screen descriptions. refer to – Hach-Lange HACH 5500 sc PO43-LR Operations User Manual

Page 6: Figure 4

background image

Figure 3 Measurement screen

1 Home (main measurement screen)

6 Relays (second icon shown if

additional relay installed)

2 Measurement channel

7 Parameter

3 Activity (shown during a

measurement or calibration
process)

8 Measurement value

4 Reminder (for a maintenance task)

9 Channel name

5 SD card (shown when a SD card is

inserted)

10 Analyzer name

Figure 4 System status screen

1 Calibration status information

3 Reagent (Rx) and standards (Sx)

with fluid level indicators (%)

1

2 PROGNOSYS service indicator bar

4 PROGNOSYS measurement

quality indicator bar

1

The number of bottles shown in the display depends on the number of bottles
installed.

PROGNOSYS indicator bars

The service indicator bar shows the number of days until a service task
is necessary. The measurement quality indicator bar shows the overall
measurement health of the analyzer measured on a scale from 0 to 100.

Color

Color meaning for service
indicator bar

Color meaning for the measurement
quality indicator bar

Green There are at least 45 days

until the next service task is
necessary.

The system is in good working
condition and the health percentage is
more than 75%.

Yellow At least one service task is

required in the next 10 to
45 days.

The system needs attention to prevent
a failure in the future. The health
percentage is between 50 and 75%.

Red

One or more service tasks are
required within the next
10 days.

The system needs immediate attention.
The health percentage is below 50%.

6 English