Option
Description
OffMax TIMER
Sets the maximum time the relay stays off (default =
0 min).
OnMin TIMER
Sets the time the relay stays on, independent from the
measured value (default = 0 min).
OffMin TIMER
Sets the time the relay stays off, independent from the
measured value (default = 0 min).
6. Select SCHEDULER in the FUNCTION menu and then select the
applicable options in the ACTIVATION menu.
Option
Description
HOLD OUTPUTS Holds or transfers outputs for the selected channels.
RUN DAYS
Selects the days the relay stays on. Options: Sunday,
Monday, Tuesday, Wednesday, Thursday, Friday,
Saturday
START TIME
Sets the start time.
INTERVAL
Sets the time between activation cycles (default = 5 min).
DURATION
Sets the period of time the relay is set to on (default =
30 sec).
OFF DELAY
Sets the time for additional hold/output time after the
relay is set to off.
7. Select WARNING in the FUNCTION menu and then select the
applicable options in the ACTIVATION menu.
Option
Description
WARNING LEVEL Sets the level for the warning activation and starts
applicable individual warnings.
Set the error hold mode
1. Select SETUP SYSTEM>SETUP OUTPUTS>ERROR HOLD MODE.
2. Select an option.
Option
Description
HOLD OUTPUTS
Holds the outputs at the last known value when
communications are lost.
TRANSFER
OUTPUTS
Switches to the transfer mode when communications
are lost. The outputs transfer to a pre-defined value.
View data
1. Select VIEW DATA.
2. Select an option.
Option
Description
ANALYZER DATA
Shows specific status information (refer to
MEASUREMENT
DATA
Shows measurement information (refer to
).
LOG DATA
Selects the data log and the event log. DATA LOG
—shows the measurement values. Select start
time, number of hours and number of readings.
EVENT LOG—shows all analyzer information.
Select start time, number of hours and number of
readings.
Table 2 Analyzer data
Element
Definition
CELL TEMP
Temperature of the colorimeter cell block heater
(ideally 49.8 °C to 50.2 °C (121.64 °F to 122.36 °F))
REAGENT TEMP
Reagent temperature before it enters the colorimeter
AMBIENT TEMP
Air temperature inside the electronics area
SAMPLE TEMP
Sample pre-heater block temperature (typically 45 °C
to 55 °C (113 °F to 131 °F) but could reach 58 °C
(136.4 °F))