beautypg.com

Calibration, Schedule automatic calibrations, Start a calibration manually – Hach-Lange HACH 5500 sc PO43-LR Operations User Manual

Page 17

background image

Note: The SD CARD SETUP option shows only when an SD card is installed.

3. Select UPGRADE SOFTWARE and confirm. Select the device and

upgrade version, if applicable.

4. When the upgrade is complete, the display shows TRANSFER

COMPLETE. Remove the SD card.

5. Restart the instrument for the upgrade to take effect.

Calibration

The automatic calibration uses the installed known standards to calibrate
the analyzer. Calibration can be done manually or scheduled to occur
automatically. Use the calibration menu to view calibration data, start or
cancel the auto calibration, override the auto cal settings, calibrate
manually or to reset to the default calibration.

1. Push cal to display the calibration menu.

Option

Description

START MANUAL CAL

Refer to

Start a calibration manually

on page 17.

SET AUTO
CALIBRATION

Refer to

Schedule automatic calibrations

on page 17.

CALIBRATION DATA

Displays data from the last calibration and the
due date and time of the next scheduled
calibration.

OVERRIDE
CALIBRATION

Enter new slope and zero values and disables
automatic calibration.

OUTPUT CALIBRATION Select a 4-20 mA output and enter the output

values to be sent.

RESET DEFAULT CAL

Reset calibration data to default values and
disables automatic calibration. On completion,
perform a new calibration.

Schedule automatic calibrations

1. Push cal, then select SET AUTO CALIBRATION.

2. Select ACTIVATE AUTO CAL>YES.
3. Select STD SOLUTION and enter the standard value in ppb.
4. Select a schedule option for calibration.

Option

Description

TIME BASE

Sets the interval between calibrations. Options: DAYS or
HOURS.

WEEK DAY

Selects the day or days of the week for calibration when
TIME BASE is set to DAY.

TIME

Sets the time of day for calibration when the TIME BASE is
set to DAY.

SET INTERVAL Sets the interval between automatic calibrations in hours

when the TIME BASE is set to HOURS.

Start a calibration manually

1. Push cal, then select START MANUAL CAL.
2. The type of calibration is displayed. Push enter to start the

calibration.

3. The measurement status is displayed. Select YES to interrupt the

current measurement cycle and start the calibration immediately.
Select NO to wait until the current measurement cycle has completed
before the calibration starts.

English 17