Power – ON/OFF
- Red indicator light.
- Normally ON when water distiller is operating.
Controls power to the entire water distiller.
- Top square indicator light is illuminated whenever the
switch is on. It will remain illuminated even when the
water distiller is not operating.
- Bottom rectangular indicator light is illuminated only
when the heating element in the boiler is on.
- Turn power switch to OFF to stop Water distiller
operation.
Fan – ON/OFF
P
O
W
E
R
(O
N
P
O
S
IT
IO
N
)
(O
N
P
O
S
IT
IO
N
)
FA
N
- Amber indicator light.
- Normally ON when water distiller is operating.
Controls power to the cooling coil fan.
- Top square indicator light is illuminated whenever the
power switch is on. It will remain illuminated even
when the water distiller is not operating.
- Bottom rectangular indicator light is illuminated only
when the fan is on. Fan should be running if the water
distiller is distilling water.
- Turn Fan Switch to OFF only when performing a
steam sterilization of the water distiller.
Overheat Reset Switch – Manual Reset
- Recessed into front of the water distiller head front
cover.
- The reset button will shut the unit off if the water
distiller has overheated, it may be reset once the water
distiller has cooled off. If the problem persists this
may also indicate a problem and should be looked at
by a service center.
- The reset button is a safety feature and is designed to
shut the water distiller down should it overheat.
- Push your finger through the hole in the front cover
labelled ‘Manual Reset’ to reset. A snap will be heard
when the switch is reset and distilling can resume.