Specifications – Flowline AXXX Smart Trak User Manual
Page 2

Step One
SPECIFICATIONS
4.2"
(106mm)
5.7"
(145mm)
2.8"
(71mm)
E (Specify)
B (Specify)
D (Specify)
C (Specify)
A (Specify)
2" NPT (2" G)
1" Square
(25.4 mm)
Specifications:
Length:
8" to 10' (20 cm to 3m)
Switch points:
1 to 4 (field adjustable)
Orientation:
± 30˚ vertical
Process temp.:
F: -40˚ to 194˚
C: -40˚ to 90˚
Pressure:
Atmospheric
Wetted material:
PP (20% glass fill)
Process mount:
2" NPT (2" G)
Enclosure rating:
NEMA 4X (IP65)
Installed height:
5.7" (14.4 cm) above tank process mount
Encl. material:
PP, UL94VO
Conduit entrance: Single, 1/2" NPT
Termination:
12 poles
Ultrasonic sensor
Supply voltage:
12-36 VDC
Consumption:
25 mA maximum
Contact type:
(1) SPST relay
Contact rating:
GP: 120 VAC/VDC @ 60 VA
IS: 32 VDC @ 0.5A
Cable type:
4-conductor / sensor, #22 AWG (shielded)
Contact output:
Selectable NO/NC
Classification:
Intrinsically safe
CE compliance:
EN 50082-2 immunity
EN 55011 emission
EN 61010-1 safety
Vibration sensor
Supply voltage:
12-36 VDC
Consumption:
25 mA maximum
Contact type:
(1) SPST relay
Contact rating:
120 VAC/VDC @ 60 VA
Cable type:
4-conductor / sensor, #22 AWG (shielded)
Contact output:
Selectable NO/NC
Classification:
General purpose
CE compliance:
EN 50082-2 immunity
EN 55011 emission
EN 61010-1 safety
Buoyancy sensor
Contact type:
(1) SPDT reed
Contact rating:
120 VAC/VDC @ 15 VA
Cable type:
3-conductor / sensor, #22 AWG (shielded)
Contact output:
Selectable NO/NC
Classification:
General purpose
CE compliance:
EN 50082-2 immunity
EN 55011 emission
EN 61010-1 safety