NOTE: damping factor will affect the display, alarms, and 4-20mA.

To turn on test mode, press “TEST” key, display will show “TEST MODE”. The 4-20mA

output will be set to 20mA. Press “ENTER” key to exit function.


To access the 2nd function, press the “2nd FUNCT” key, display will show “SERVICE

MODE”. Then select any function that is located in the yellow part of the key pad.

Press “ENTER” key to exit function.


To view the Doppler’s signal strength level, press “2nd FUNCT” key, then “SIGNAL

STR” key. The display will show “SERVICE MODE” and “SIGNAL STR.”. When flow is

present, a good signal strength level is 150 or higher. Press ‘ENTER’ key to exit

function.


To adjust the 4mA output press “2nd FUNCT” key then the “4mA” key. The

µP display

will show “SERVICE MODE” and “4mA”. Use the ‘SCROLL” keys to increase or

decrease the 4-20mA output level. When finished with function, press “ENTER” key.


To view the Doppler’s raw velocity level, press “2nd FUNCT” key and then “VEL ADC”

key. The

µP display will show “SERVICE MODE” and “VEL ADC1”. When flow is at 10

FPS, a level of 1024 will be displayed. Press “ENTER” key to exit function.


To set the liquid Blanking time, press “2nd FUNCT” key and then the “BLANKING” key.

The

µP display will show “SERVICE MODE” and “BLANKING”. Use the “SCROLL”

keys to increase or decease the minimum liquid level the unit can read. This is done by

rejecting receive signals that are less than the set value 17 (0” -2.0”) or (1 blanking

change is equal to 0.2” liquid level). Changing the Blanking setting may also alleviate

unstable readings in some applications. Default level is 20

[0” - 2.6”]. Press “ENTER”

key to exit function.

NOTE: never set the blanking range greater than 75% of the minimum level you desire to detect.