beautypg.com

Dynasonics PFP-600 User Manual

Page 14

background image

units = 1 & 9
totalizer is off and no display

units = 5 & 10
totalizer is off and no display



To set totalizer multiplier, press “TOT MULT” key. The

µP display will show “TOTAL

MULT.” Use the “SCROLL” keys to increase or decrease value. The default value is 0.

When finished, press “ENTER” key to exit.

0 = X1
1 = X10
2 = X100
3 = X1000
4 = X10,000


To set totalizer on, off or reset, press “TOT ON/OFF” key once to stop totalizing, twice

to reset, and third to start. Note: when changing engineering units, totalizer will stop

totalizing.


To enable key board lock out, press “LOCK ON” key, display will show “LOCK ON”,

then press “ENTER” key to exit. To turn off lock, use arrows to set a value of 125, and

press “ENTER” key, display will show “LOCK OFF”.

“F1” & “F2” keys are not used.


To set velocity calibration factor, press “CAL” key. The

µP display will show “VEL.

CAL”. Use the “SCROLL” keys to increase or decrease value, cal. range is 0-255%,

default is 100%. When finished, press “ENTER” key to exit. NOTE: calibration factor

will affect the display, alarms, and 4-20mA.


To set damping factor, press “DAMP” key. The

µP display will show “DAMPING”. Use

the “SCROLL” keys to increase or decrease value, damping range is 1-10 seconds,

default is 2 seconds. When finished, press “ENTER” key to exit.

12