Dynasonics PFP-600 User Manual
Page 14

units = 1 & 9
totalizer is off and no display
units = 5 & 10
totalizer is off and no display
To set totalizer multiplier, press “TOT MULT” key. The
µP display will show “TOTAL
MULT.” Use the “SCROLL” keys to increase or decrease value. The default value is 0.
When finished, press “ENTER” key to exit.
0 = X1
1 = X10
2 = X100
3 = X1000
4 = X10,000
To set totalizer on, off or reset, press “TOT ON/OFF” key once to stop totalizing, twice
to reset, and third to start. Note: when changing engineering units, totalizer will stop
totalizing.
To enable key board lock out, press “LOCK ON” key, display will show “LOCK ON”,
then press “ENTER” key to exit. To turn off lock, use arrows to set a value of 125, and
press “ENTER” key, display will show “LOCK OFF”.
“F1” & “F2” keys are not used.
To set velocity calibration factor, press “CAL” key. The
µP display will show “VEL.
CAL”. Use the “SCROLL” keys to increase or decrease value, cal. range is 0-255%,
default is 100%. When finished, press “ENTER” key to exit. NOTE: calibration factor
will affect the display, alarms, and 4-20mA.
To set damping factor, press “DAMP” key. The
µP display will show “DAMPING”. Use
the “SCROLL” keys to increase or decrease value, damping range is 1-10 seconds,
default is 2 seconds. When finished, press “ENTER” key to exit.
12