Refer to the following table if your Glassware Washer fails to operate properly. If
the suggested corrective actions do not solve your problem, contact Labconco
Product Service for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Detergent cup fails
to open
Malfunction of cup
Replace cup.
Malfunction of relay
Check electrical continuity or replace PCB.
Glassware interfering
with cup door
Move glassware.
Noises during
operation
Glassware rattling
Load glassware properly. Do not allow glassware
to touch.
Door panel vibrating
Tighten panel screws.
Wash arms hitting
Check glassware loading.
Water inlet line
makes occasional
hammering sound
Check with plumber.
Purified water pump
is activated
No corrective action needed. This is normal for
about 5 minutes while pure water enters the
washer.
Door drops too far
(below door stop)
Bent door hinge
Replace door hinge.