Carolina 140 Series General Purpose Incubators User Manual
Specifications, Common unit construction, Common unit specifications

PAGE 1
Issue Rev A 1208
Quincy Lab Inc. 1925 N. Leamington Ave. Chicago, Illinois 60639 1-800-482-HEAT (4328)
Copyright Quincy Lab Inc. 2008
Model Series 140 & 180
General Purpose Incubators
OPERATING MANUAL
Quincy Lab, Inc.
HEAT/CYC
LE
POWER
Model 12-140 Incubator
Interior Dimensions
INCHES W x H x D
(CM) W x H x D
Exterior Dimensions
INCHES W x H x D
(CM) W x H x D
12x10x10
18x16x12
18x16x12
12x10x10
13x15x11
19x21x13
13x15x11
19x21x11
Weight (lbs)
19 lbs
33 lbs
19 lbs
33 lbs
Standard Electrical
31x25x25
46x41x30
46x41x30
31x25x25
33x38x28
48x53x33
33x38x28
48x53x33
SPECIFICATIONS
MODEL
10-140
MODEL
12-140
MODEL
10-180
MODEL
12-180
Temperature Range
Ambient + 2
o
C to 62
o
C
Common Unit Construction
Exterior: Powder-Coated Steel Interior: Aluminum
Insulation: Fiberglass
Door: 140: Acrylic, 180: Steel Insulated
Thermo-control: Bi-Metal
Heater: Resistive-Tubular Incoloy
* Standard models voltage only, optional 230 voltage available. Check label on back of unit.
VOLTS / WATTS
115 / 120*
115 / 235*
115 / 270*
115 / 385*
HEAT/CYCLE
POWER
Model 10-180
Incubator
Ambient + 3
o
C to 94
o
C
Cubic Foot Capacity
.7 ft
3
2.0 ft
3
.7 ft
3
2.0 ft
3
Model 10-180
Model 12-140
Common Unit Specifications
Operating Environment: Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,
Pollution Degree 2, ambient temperature 10
o
C/50
o
F to 35
o
C/95
o
F,
80% RH maximum.
Storage Temperature: -10
o
C/14
o
F to 70
o
C/158
o
F, 70% RH maximum.
Approvals:
Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)
Compliance: UL Standard 61010-1, IEC 61010-1, 2nd Edition.