NEULOG OXYGEN SENSOR GUIDE

Operation with NeuLog viewer

Prior to initial use the Neulog oxygen sensor needs to be calibrated, please review the
“Sensor Calibration” section of this document for detailed instructions.

Materials needed:

NUL-205 Oxygen Sensor

VIEW-101 Viewing Module

BAT-200 Battery


Procedure:

1.

Calibrate the oxygen sensor. (Follow the „Sensor Calibration” guide in this
document)

2.

Connect the oxygen sensor to the left side of the viewer.

3.

Connect the BAT-200 to the right side of the viewer.

4.

The VIEW-101 will auto detect the sensor and you will see it appear on the
left side of the screen.

5.

Once the sensor appears it will be monitoring data in real time.

6.

Remove the blue rubber cap on the end of the oxygen probe.

7.

To run an experiment and collect data click the run button (little green
person)


Off-line experiments

(Off-line experiments are for when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).

Materials needed:

NUL-205 Oxygen Sensor

BAT-200 Battery


Materials needed to configure your offline experiment:

USB-200 USB Module

or

WIFI-201 WIFI module

or

VIEW-101 Viewing Module

 A USB to mini USB cable (which comes with the USB-200)

Procedure:

1.

Calibrate the oxygen sensor. (Follow the „Sensor Calibration” guide in this
document.

2.

Connect oxygen sensor directly to the left side of the charged BAT-200
Module

3.

Remove the blue rubber cap on the end of the oxygen probe.

4.

When ready to collect data press the “Start/Stop” button on the oxygen
Sensor, a red light will turn on for the duration of the experiment.

5.

To change the experiment settings first connect to your chosen device (PC,
Mac, tablet, smart device) and select the “Module setup” button. (For more
information on how to change experiment settings view quick start
procedure section for your chosen device in this document.)

6.

After the experiment has concluded (when the “Start/Stop” button is pressed
again or the data collection period ends) connect the sensor to your chosen
device. (full instructions in each device‟s quick start procedure section)

7.

Open Neulog software.

8.

Click the “Off-line experiment” button.

9.

Click the “Load data from sensors” button.

10. Select which experimental data to upload. (5 experimental runs can be

stored at one time on the NUL-205 oxygen Sensor)


Restoring sensor’s factory default settings:


Procedure:

1.

Connect the USB-200 to a PC, Mac, tablet, or smart device.

2.

Connect the oxygen sensor to a USB-200 module (they directly plug
together).

3.

Open the NeuLog software.

4.

Click the “Tools” icon.

5.

Click “Restore sensor‟s factory defaults” icon.

6.

If prompted to clear the graph, click either the “Off-line experiment” or “On-
line experiment” (whichever you are using).

7.

Click “Clear experiment results”.

8.

Resume from step 4.