OVERVIEW
The Schreiber Medical Chillers are designed to supply constant chilled water to an MRI
system or other scanning system that requires chilled water. These chillers are also
designed for Outdoor or Indoor installation. However, if indoor installation is desired,
adequate ventilation is required. If necessary, consult with Schreiber Engineering for
more information.
Each chiller is self contained and includes 2 pumps, 2 independent refrigeration
modules; air cooled condenser and fans, control system, and a chilled water reservoir.
The modular construction allows for quick and easy maintenance service.
Schreiber Engineering Medical Chillers utilize microprocessor controller via remote
terminal allowing monitoring the chiller operation at two (2) locations. The 4 x 20 LCD
remote terminal shows system status, alarms and diagnostics procedures.
If required, the chiller can be supplied in a Remote Condensing (RC) version. This is for
installations where the chiller is mounted indoors and a remote condenser is installed
outdoors.
A QUICK ELECTRICAL SPECIFICATION:
VOLTAGE
FLA
MIN. SUPPLY
MAX FUSE
460/3 40.15 48.8 55
230/3 78.2 95
110