INSTALLATION INSTRUCTIONS

Air or Water Cooled Chillers

1. When the chiller is received, check for any shipping damage. If any is found, a

freight claim should be filed at once with the carrier.

2. This water chiller is designed with a contaminant free chilled water system,

stainless steel reservoir, and all copper evaporator. Installed properly, it will give

you years of trouble free service. Therefore, it is important that the chilled water

piping system you are using to plumb this water chiller is also as contaminant free

as possible. When you take this into consideration you will have a most efficient

water cooling system

Should you need assistance to clean up an old chilled water system or to keep a

new system clean, we suggest you contact your local water treatment company.

3. Place the water chiller in an area that allows easy access to the service doors and

control panel for routine maintenance and inspection. For the air-cooled unit, make

sure it is in an area where there is good air circulation over the condenser inlet,

24"

from any obstruction. For the water-cooled unit, make sure it has a good water

supply by keeping the piping to the condenser 1" in size. Restriction of air to the

air-cooled condenser or water to the water-cooled unit will not only affect the

efficiency of the unit, but will also cause damage to the refrigeration system. Note:

Do not mix up the inlet and the outlet to the condenser on water-cooled units.

Water-cooled takes 3 gpm, per ton.

4. Make sure the ceiling above the fan discharge is over 12 feet to avoid recirculation

of the condenser air, because recirculating condensed air can cause the unit to cut

off on the high-pressure switch. If the ceiling is lower than 12 feet, a straight duct

out of the ceiling should be installed in order to prevent the air from recirculating

over the condenser.

5. Make up two chilled water lines, a discharge and a return line. The fitting on the

chiller for these lines is 1" pipe coupling. IT SHOULD BE NOTED THAT THERE IS

A 1/4" COPPER BY-PASS LINE BETWEEN THE WATER PUMP AND THE

WATER TANK. This by-pass allows just enough water to circulate through the

water system in order to prevent damage to the water pump in case the chilled

water discharge is shut off. This small amount of water will not affect the efficiency

of the unit.

6. Fill the chilled water tank through the hose bib, which is located on the chilled water

return line. Fill the tank until it reaches the level as marked on the water sight glass.

The water level should be kept within 1" of the marked level.

Do not fill the water

tank with deionized water.