Step 9

– Repeat step configurations

Repeat steps 1-8, except in step one,
select step number as an increment
to the next step. Create a sequence
of 5 steps with the following
parameters:

Steps

Type

Data Value Delay Time

Upper

Voltage

Limit

Lower

Voltage

Limit

Upper

Current

Limit

Lower

Current

Limit

1

CV

2.000 V

5 s

60.00 V

0.00 V

30.00 A

0.00 A

2

CR

10.00

5 s

60.00 V

0.00 V

30.00 A

0.00 A

3

CC

2.000 A

5 s

60.00 V

0.00 V

30.00 A

0.00 A

4

CC

0.000 A

5 s

60.00 V

0.00 V

30.00 A

0.00 A

NOTE

Steps 4 has been set with data value of 0.000 A to denote the end of the
sequence. Always set the last step of your sequence to 0.000 under data value
if the sequence has less than 25 steps. Otherwise, the load will only stop at the
step that contains either a data value of 0 or after the 25

th

step.


Step 10

– Save and Exit LIST menu.

Once finished, press the

Shift

key,

then press

C

View

key to save

and exit the programmed
sequence(s).

NOTE

Users can program all 100 steps (0 to 99) at once without having to select the
memory storage group.

Following the example, the below sequence is programmed and can be
ran from the instrument.

10

Ω, CR




1 A, CC

2 V, CV

2 A, CC