3 general errors, General errors – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual
Page 66

What to do if...
pH/ 3400i
66
Measured value too low
Measured value too
high
6.3
General errors
Display LoBat
Instrument does not
react
to keystroke
Display
to
to
to
to
Probe symbol flashes
Message
StoFull
StoFull
StoFull
StoFull
Cause
Remedy
– Insufficient flow
– Provide flow to the probe
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function
using the salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the
membrane
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
– Batteries almost empty
– Exchange the batteries (see
section 5.1 M
AINTENANCE
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Switch the instrument on
while pressing the
Cause
Remedy
– Time-out of the interface
– Check the instrument that is
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– All memory locations are full
– Output data storage and clear
data storage