pH/ 3400i

Operation

37

4.4.5

D. O. calibration

Why calibrate?

D. O. probes age. This changes the slope of the D. O. probe. Calibra-
tion determines the current slope of the probe and stores this value in
the instrument.

When to calibrate?

z After connecting another D. O. probe
z When the probe symbol flashes (after the calibration interval has ex-

pired).

Calibration procedures

The calibration is performed in water vapor-saturated air. Use the
OxiCal

®

-SL air calibration vessel (accessories) for the calibration.

AutoRead

The calibration procedure automatically activates the AutoRead func-
tion. The AR display indicator flashes. The calibration process is fin-
ished when AR stops flashing.

Displaying the calibra-

tion data

You can view the data of the last calibration on the display. The pro-
ceeding is described on page 48.

Printing the

calibration protocol

The calibration record contains the calibration data of the current cali-
bration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 50).

Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.7.2
before calibrating. After a valid calibration, the record is printed.