beautypg.com

9 reset, Reset – Xylem MULTI 3400i pH/Oxi MULTI-PARAMETER METERS User Manual

Page 58

background image

Operation

pH/ 3400i

58

4.9

Reset

You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.

Measurement

parameters

The following measuring parameters are reset to the delivery status:

pH

Oxi

Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery
status:

Measuring mode

pH

Slope -59.16 mV/pH

Asymmetry

0 mV

Manual temperature input

25 °C

Measuring mode

D. O. concentration

Relative slope

1.00

Salinity (value)

0.0

Salinity (function)

off

Baud rate

4800

Display during the
pH calibration

Buffer nominal value

Interval 1
(automatic save)

OFF

Interval 2
(for data transmission)

OFF