Oxi 1970i

What to do if...

49

ba75691e01

01/2007

6

What to do if...

Display LoBat

Instrument does not

react to keystroke

Error message

0FL

0FL

0FL

0FL

Error message

E3

E3

E3

E3

Cause

Remedy

– Battery almost empty

– Charge the battery (see

section 3.2)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the and
keys at the same
time and release them again.
The software version is
displayed.

Cause

Remedy

Display range exceeded

Oxygen sensor:

– Not connected

– Connect the sensor

– Cable broken

– Replace sensor

– Depleted

– Replace sensor

– Short-circuit between gold and

lead electrode

– Clean sensor and replace it if

necessary

Cause

Remedy

Invalid calibration

Oxygen sensor:

– Electrolyte solution depleted

– Regenerate sensor

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate sensor

– Obsolete

– Replace sensor

– Broken

– Replace sensor