1 calibrating with the air calibration vessel, Calibrating with the air calibration vessel – Xylem Oxi 1970i PORTABLE DISSOLVED OXYGEN METERS User Manual

Page 23

background image

Oxi 1970i

Operation

23

ba75691e01

01/2007

4.3.1

Calibrating with the air calibration vessel

Proceed as follows to calibrate the instrument:

Note
The sponge in the air calibration vessel must be moist (not wet). Follow
the instructions in the OxiCal

®

-SL or DurOx 325 operating manual.

1

Connect the D. O. sensor to the measuring instrument.

2

Keep an air calibration vessel ready for use.

3

Put the D. O. sensor into the air calibration vessel.

4

Press the key repeatedly until O

2

CAL appears.

5

Press the key. AutoRead is active, AR flashes.

6

As soon as a stable value is achieved, AR stops flashing. Cal-
ibration is finished; the sensor symbol indicates the relative
slope determined and the sensor evaluation (see page 22).