5 transmitting data, 1 data transmission interval (int 2), Transmitting data 4.5.1 – Xylem Oxi 1970i PORTABLE DISSOLVED OXYGEN METERS User Manual
Page 37: Data transmission interval (int 2)

Oxi 1970i
Operation
37
ba75691e01
01/2007
4.5
Transmitting data
You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 30), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured
values are periodically output on the interface (see below).
z With the Output data storage function (page 32), calibration data or
saved measured values are output on the interface.
z Via the analog recorder output (page 39), measured values are out-
put as voltage values.
z With the KOM pilot communication kit (accessory), data can be
transmitted bidirectionally (page 40).
Note
If you connect a recorder (analog output), the output on the digital in-
terface is switched off.
4.5.1
Data transmission interval (Int 2)
The interval for the data transmission (Int 2) determines the chronolog-
ical interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
Note
When the AutoStore function is active, the data transmission is per-
formed according to the setting of the save interval (Int 1). Set the save
interval (Int 1) to OFF to activate the Data transmission interval (Int 2).