XVIII.

Maintenance:

a. Except for battery and desiccant, there are no user-serviceable parts

inside the WL16 Data Logger.

b. Global Water recommends checking the batteries every 6 months.
c. Clean the sensor screen using clean water and a toothbrush to

remove any algae, mud or sludge.

XIX. Troubleshooting:

Issue: Logger is reading incorrectly.

a. Verify the battery supply has enough voltage. To operate properly, the

logger and sensor must have a minimum of 8 volts. If the battery

voltage monitor is reading less than 15 volts or seems incorrect, check

the batteries with a voltmeter or replace them. Battery voltages of less

than 15 volts indicate that the batteries need replacing.

b. Check the logger and sensor calibration numbers in the Setup Menu.

Recalibrate the logger if necessary. As a precaution, save the setup

file first as described in section XI. Clean the sensor screen.

Issue: Cannot communicate with the logger.

a. Verify that good batteries are installed.

b. Check the communication cable.

c. If the connection is to the WL16S using the serial cable provided by

Global Water, verify that the correct COM port is selected in the

connection menu of the Global Logger software.

d. If the connection is to the WL16U through the USB port, use the Device

Manager to confirm that the driver is loaded and the correct virtual

COM port is selected in the connection menu of the Global Logger

software. Check the Device Manager in your operating system and

confirm that the USB COM port is not being used by another device.

e. Try reducing the baud rate in the connection screen of the Global

Logger software.