6.2
ISE system messages
Error message
0FL
0FL
0FL
0FL
Error message
E3
E3
E3
E3
Concentration
display flashes
Cause
Remedy
Ion-selective electrode:
– Not connected
– Connect the electrode
– Cable broken
– Replace the electrode
Cause
Remedy
Ion-selective electrode:
– Slope not in the range
50 ... 70 mV or
23 ... 35 mV
– Replace the electrode
– recalibrate
– Moisture in the plug
– Dry plug
– Electrode obsolete
– Replace the electrode
– Electrode broken
– Replace the electrode
Measuring instrument:
– Calibration procedures:
Incorrect order of the
standards with 3-point
calibration
– Select the correct order
– Socket damp
– Dry socket
Cause
Remedy
In measurement:
– The measured value is outside
the range determined through
calibration
– Select the calibration
standards so that the
measured value is in between
them
In calibration:
– Calibration standards lie too
closely together. Select other
calibration standards
U2
U1
–
5mV
≤
U3
U2
–
5mV
≤
- Lab 850 BENCHTOP pH METER (73 pages)
- Lab 850, up to serial number 09470314 (61 pages)
- Lab 860 BENCHTOP pH METER (74 pages)
- Lab 870 BENCHTOP pH METER (71 pages)
- Lab 960 (77 pages)
- Lab 970 (71 pages)
- CG 839 (8 pages)
- CG 842 (44 pages)
- CG 843P (72 pages)
- CG 843 (68 pages)
- CG 853P (76 pages)
- CG 853 (74 pages)
- CG 855 (11 pages)
- HandyLab 100 QuickStart (50 pages)
- HandyLab 100 (66 pages)
- HandyLab 200 QuickStart (50 pages)
- HandyLab 200 (52 pages)
- HandyLab 600 QuickStart (50 pages)
- HandyLab 600 (72 pages)
- HandyLab 680 QuickStart (50 pages)
- HandyLab 680 (120 pages)
- HandyLab 700(EX) (128 pages)
- HandyLab 700 (76 pages)
- HandyLab 780 (164 pages)
- handylab LF 11 (44 pages)
- handylab LF 12 (64 pages)
- handylab OX 12 (60 pages)
- handylab multi 12 (82 pages)
- handylab pH 11 (48 pages)
- handylab pH 12 (68 pages)
- handylab pH LF 12 (80 pages)
- IQ SensorNet FDO 70x IQ (SW) (48 pages)
- IQ SensorNet ViSolid 700 IQ (SW) (58 pages)
- MultiLab 4010-2 (112 pages)
- MultiLab 4010-3 (112 pages)
- EBI 40 TK-12 (18 pages)
- TFX 410 (38 pages)
- EBI 20 Serie (20 pages)
- EBI20 TEMPERATURE/HUMIDITY DATALOGGERS (40 pages)
- EBI 20-TF (40 pages)
- TFX 422 (36 pages)
- ST 1000 (42 pages)
- TLC 1598 (2 pages)
- TTX 100/110/120 (28 pages)
- EBI 25 (18 pages)