2c d – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 27

pH/ 3400i
Operation
27
Note
At this point, the AutoCal DIN calibration can be terminated with
This corresponds to a single-point calibration. When doing so, the in-
strument uses the Nernst slope (-59,16 mV/pH at 25 °C) and deter-
mines the asymmetry of the electrode.
5
When the measured value is stable, Cd2 appears.
6
Thoroughly rinse the electrode with deionized water.
7
If required, enter the temperature of the second buffer solution
with <
▲> <▼> while keeping the
pressed.
8
Immerse the pH electrode in the second buffer solution.
9
Press the
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
AutoCal TEC
6
mV/pH
mol/L
8
Auto Store
RCL
LoBat
CalError
ARng AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2
C d
°C
TP
pH
AutoCal DIN
ISE
U
AutoCal TEC
AutoCal TEC
mV
U
mV/pH
mol/L
8
Auto Store
TempError
RCL
LoBat
CalError
ARng
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
%
°F
S
mg/L
% ppm
8
4
2
°C
TP
mV/pH
mol/L
8
Auto Store
TempError
RCL
LoBat
CalError
ARng
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
%
°F
S
mg/L
% ppm
ISE
8
4
2
0
8
°C
TP
2 3
7
pH
mV
AR
1
AR
AutoCal DIN
AutoCal DIN
pH
1
9