H-4280

SDI-12 Command and Response Protocol 3-13

3.10 Extended Set_Current_Stage Command

The H-4280 processes each voltage measurement with a y=mX+b linear equation. During

installation it is often convenient to quickly set the H-4280's measurement output to a certain

value. For example, the output can be quickly set to match the current stage of a water elevation

sensor as determined by a staff gauge or other datum. The “XSCS” extended command causes

the H-4280 to make a fresh measurement and automatically update the Offset (b) term as needed

to produce the desired value.

Command

"aXSCSn ddd!"

Where: a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XSCS

are upper case characters

n

is the channel number (0-7)

ddd

is the new stage setting

(Note: there must be a space between the n and ddd arguments)

This command places 1 value in the data buffer. Use the “aD0" command to collect and view the

current value.

Example of a H-4280 Extended "Set Current Stage" command:

Command

Response

Time Values

Description

"aXSCS3 2.34!"

"a0031"

3 sec 1

Set the Stage to 2.34

Subsequent Command

Response

Description

"aD0"

a+12.80

The new Offset for channel 3