3-4 SDI-12 Command and Response Protocol
H-4280
3.3 Concurrent Measurement Command
This is a new command for the Version 1.2 SDI-12 Specification. A concurrent measurement is
one which occurs while other SDI-12 sensors on the bus are also taking measurements. This
command is similar to the “aM!” command, however, the nn field has an extra digit and the sensor
does not issue a service request when it has completed the measurement. Communicating with
other sensors will NOT abort a concurrent measurement. Data values generated in response to
this command are stored in the sensor's buffer for subsequent collection using "D" commands. The
data will be retained in the sensor until another "M", "C", or "V" command is executed.
Command
Response
Description
"aC!"
"atttnn
Initiate measurement
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
C
is an upper-case ASCII character
ttt
is a three digit integer (000-999) specifying the maximum time, in seconds, the
sensor will take to complete the command and have measurement data available in
its buffer.
nn
is a two digit integer (00-99) specifying the number of values that will be
placed in the data buffer. If "n" is zero (0), no data will be available using
subsequent "D" commands.
The data recorder may wake the sensor with a break and collect the data anytime after the
specified processing time has elapsed.