beautypg.com

H-4280 – Xylem H-4280 User Manual

Page 19

background image

H-4280

SDI-12 Command and Response Protocol 3-7

3.6 Initiate Verify Command

The Verify Command causes a verify sequence to be performed. The result of this command is similar to

the "aM!" command except that the values generated are fixed test data and the results of diagnostic

checksum tests. The data generated in response to this command is placed in the sensor's buffer for

subsequent collection using "D" commands. The data will be retained in the sensor until another "M", "C",

or "V" command is executed.

Command

Response

Description

"aV!"

"atttn"

Initiate verify sequence

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

V

is an upper-case ASCII character.

ttt

is a three digit integer (000-999) specifying the maximum time, in seconds, the sensor will

take to complete the command and have data available in its buffer.

n

is a single digit integer (0-9) specifying the number of values that will be

placed in the data buffer. If "n" is zero (0), no data will be available using

subsequent "D" commands

Example of a "aV!" command:

Command

Response

Time Values Description

"aV!"

"a0013"

1 sec

3

Return fixed data and diagnostic data for testing

purposes.

Subsequent Command

Response

"aD0"

a+123.456+78.9+y

Key

Description

Units

+123.456

Fixed test data

+78.9

Fixed test data

y

ROM checksum test

0 = Failed, 1 = Passed