beautypg.com

Care, maintenance and storage, General maintenance, Sensor maintenance – YSI Pro30 User Manual

Page 13: General maintenance sensor maintenance

background image

18

19

Figure 10, Erase data mode.

Use the up or down arrow key to select Erase Data Set, Erase All Sets or the ESC-

Exit function box, then press enter to confirm.

Select ESC-Exit and press enter to exit Erase mode without erasing any data.

Select Erase Data Set and press enter to erase the data set that was displayed

before entering Erase mode. For example, if data set 12 was displayed before

entering erase mode, and Erase Data Set is selected, Data Set 12 will be erased

from memory and the data sets AFTER that number will move up to keep them

sequential. For example, if there are15 records and number 12 is erased then 13

becomes 12, 14 becomes 13, and 15 becomes 14. The instrument will return to

Data mode after erasing one data set.

Select Erase All Data Sets and press enter to clear the Pro30 memory and return

to Data mode.

EXITING DATA MoDE

While in Data mode, press enter to access the bottom functions. Next, highlight

the ESC-Exit box and press enter to return to the Run screen.

C A R E , M A I N T E N A N C E A N D

S T o R A G E

This section describes the proper procedures for care, maintenance and storage

of the instrument. The goal is to maximize their lifetime and minimize down-

time associated with improper instrument usage.

general maIntenance
GENERAL MAINTENANCE - GASKET

The instrument utilizes a gasket as a seal to prevent water from entering the

battery compartment. Following the recommended procedures will help keep

the instrument functioning properly.

If the gasket and sealing surfaces are not maintained properly, it is possible that

water can enter the battery compartment. If water enters this area, it can severely

damage the battery terminals causing loss of battery power and corrosion to the

battery terminals. Therefore, when the battery compartment lid is removed, the

gasket that provides the seal should be carefully inspected for contamination (i.e.

debris, grit, etc.) and cleaned with water and mild detergent if necessary.

sensor maIntenance
SENSoR MAINTENANCE - TEMPERATURE

You must keep the temperature sensor free of build up. other than that, no

additional maintenance is required. A toothbrush can be used to scrub the

temperature sensor if needed.

SENSoR MAINTENANCE - CoNDUCTIVITY

The openings that allow sample access to the conductivity electrodes should be

cleaned regularly. The small cleaning brush included in the Maintenance Kit is

intended for this purpose. Dip the brush in clean water and insert it into each

hole 10 to 12 times. In the event that deposits have formed on the electrodes, it

may be necessary to use a mild detergent (laboratory grade soap or bathroom

foaming tile cleaner) with the brush. Rinse thoroughly with clean water, then

check the response and accuracy of the conductivity cell with a calibration

solution.