9 manual control, 10 alarm contact, See section 4.5.9 – YSI IQ SensorNet MIQ/R6 Module User Manual
Page 41: See section 4.5.10

MIQ/R6
Settings
4 - 27
ba76031e01
01/2012
4.5.9
Manual control
Function
With the Manual control function, you can test the operativeness of an
instrument that is connected to the relay. To do so, close or open the
relay manually and, while doing so, check the behavior of the
connected instrument.
Note
The settings for other functions in the Relay function menu, as for
example Frequency controller and Pulse-width output, are retained
while the Manual control is carried out.
4.5.10 Alarm contact
Function
The Alarm contact function triggers a relay action (close or open) if a
specified alarm event occurs. The Alarm contact function is only
available for relays that are not linked with a sensor. If necessary, an
existing link has to be deleted.
Note
How to set up alarms is described in the IQ S
ENSOR
N
ET
system
operating manual of the system 2020 XT. Other possibilities of how to
use the alarm function are described there as well.
Setting
Selection/Values
Explanation
Relay function
Manual control
The selected relay action
is carried out with Save
and quit.
Action
Open
Close
Relay action (see
section 4.5.1)
Setting
Selection/Values
Explanation
Relay function
Alarm contact
The selected relay action
is carried out with Save
and quit.
Action
Open
Close
Relay action (see
section 4.5.1)