Warning, Connecting lines to the terminal strip, Miq/r6 installation 3 - 5 – YSI IQ SensorNet MIQ/R6 Module User Manual
Page 13: Fig. 3-1 terminal strip with the relay connections, Ensor, Relay output, Segment 2 segment 1

MIQ/R6
Installation
3 - 5
ba76031e01
01/2012
Connecting lines to the
terminal strip
Fig. 3-1
Terminal strip with the relay connections
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that safe areas could come into
contact with dangerous voltages. This could result in life
threatening electric shock when working with the IQ S
ENSOR
N
ET
.
Always cut off any wires that are not in use as closely as possible
to the cable gland.
1
Open the module.
2
Screw the cable gland (pos. 1 in Fig. 3-1) with the sealing ring
(pos. 2) into the module housing.
3
Loosen the coupling ring (pos. 3 in Fig. 3-1).
4
Feed the line through the cable gland in the module housing.
5
Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the label located under the
terminal strip.
6
Tighten the coupling ring (pos. 3 in Fig. 3-1).